Microscope Failure Causes
Edit: Shanghai New Eyes Medical Inc    Date: May 07, 2016

A, and verification method put standard carved line feet placed in hardness meter (or microscope) of work stage, check Shi first adjustable good focal length, makes in eyepiece vision within or projection screen Shang can clear to see standard carved line feet of carved line, and adjustment to and eyepiece within of carved line coincide, then will readings microscope of carved line and standard carved line feet of carved line for compared, should at least in whole measurement range of 5 a interval paragraph for measurement, the interval paragraph compared 3 times, take 3 times compared results of average, its relative errors w by Xia type for calculation:

W=(Li-L)/L×100%

Type: error W--(mm), Li--a reading microscope comparison of the measured length(mm), (i=1~5), L--standard reticle scale comparison of actual length (mm).

Reading microscope scale compares paragraph by paragraph as described above, the error shall not exceed ± 0.5%.

Second, failure causes and adjustment

1. microscope cloudy clear

The main reason: lens dirty or moldy.

Troubleshooting: when there is dust or dirt on the lens, using brush, feather drop, then with lens paper or with cotton wool dipped in alcohol or ether, wipe carefully along a circular path, but do not wipe the fluid loss.

2. can't clearly see the indentation in the mirror edge

The main reason: some loosening of the lens.

Elimination: the fastening of lens is loose again.

3. a reading microscope scale value and the gauge scale do not coincide

The main reason: lens lenses loose lens or objective lens tube gasket at the connection is lost, the focal length changes.

Elimination: the objective lens fastening, if washer is missing, and after fiddling withthe thickness, with suitable washers, to minimum position calibration error.

4. a reading microscope scale value greater than the gauge scale

The main reason: lens barrel increase, may be loose tube joints.

Troubleshooting: retighten tube connector.

A reading microscope should be kept clean, long time, can be placed in the dryer toprevent mold. Use the process to gently, so as not to damage the microscope, affect measurement accuracy and service life.